FDA Adverse Event Malfunction Summary report: N

RELIANCE T HANDLE

MDR report key: 3970189 · Received July 31, 2014

Report

Report Number
0009617544-2014-00342
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
STRYKER SPINE-FRANCE
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD: VISUAL INSPECTION; FUNCTIONAL INSPECTION; DEVICE HISTORY REVIEW; COMPLAINT HISTORY REVIEW; RISK ASSESSMENT; RESULTS: NO RELEVANT ISSUES WERE FOUND IN THE MANUFACTURING RECORDS OF THE DEVICE LOT THAT MAY HAVE CONTRIBUTED TO THE REPORTED EVENT. THE IFU FOR INSTRUMENTS STATES THE INSTRUMENTS MAY FRACTURE DEPENDING ON THE NUMBER OF TIMES THEY ARE USED AS WELL AS THE PRECAUTIONS TAKEN IN HANDLING, CLEANING AND STORAGE. ALSO, INSTRUMENTS WITH ARTICULATIONS MAY REQUIRE LUBRICATION. THE T-HANDLE FRACTURED AS A RESULT OF MIXED CLEAVAGE AND DUCTILE OVERLOAD. CONCLUSION: THE PLAUSIBLE ROOT CAUSE OF THE REPORTED EVENT IS USER RELATED, SPECIFICALLY, APPLYING EXCESSIVE FORCE TO THE T-HANDLE DURING ASSEMBLY.

Description of Event or Problem · 1

IT¿S REPORTED THAT DURING SURGERY, THE SURGEON USED THE REAMER DISTRACTOR (10MM). THE REAMER DISTRACTOR WAS ASSEMBLED TO THE T-HANDLE. THE SURGEON WAS NOT ABLE TO REMOVE REAMER DISTRACTOR FROM T-HANDLE AFTER USE. ALTHOUGH THE SURGEON REMOVED REAMER DISTRACTOR FORCIBLY, THE PARTS INSIDE T-HANDLE BROKE. THE SURGEON CHANGED THE T-HANDLE TO ANOTHER T-HANDLE.

Description of Event or Problem · 1

IT铠REPORTED THAT DURING SURGERY, THE SURGEON USED THE REAMER DISTRACTOR (10MM). THE REAMER DISTRACTOR WAS ASSEMBLED TO THE T-HANDLE. THE SURGEON WAS NOT ABLE TO REMOVE REAMER DISTRACTOR FROM T-HANDLE AFTER USE. ALTHOUGH THE SURGEON REMOVED REAMER DISTRACTOR FORCIBLY, THE PARTS INSIDE T-HANDLE BROKE. THE SURGEON CHANGED THE T-HANDLE TO ANOTHER T-HANDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
448378 RELIANCE T HANDLE INSTRUMENT-T-HANDLE LXH STRYKER SPINE-FRANCE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1