ES6 STERNUM SAW
Report
- Report Number
- 0001811755-2014-02731
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 4, 2014
- Report Date
- July 4, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- OTHER
Narratives
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
THE REPORTED EVENT WAS CONFIRMED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH FUNCTIONAL EVALUATION. UPON DISASSEMBLY, A FAILED POTTED TRIGGER ASSEMBLY WAS FOUND, WHICH CAN LEAD TO THE REPORTED EVENT.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY, THE DEVICE CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION, AND SUBSEQUENTLY THE DEVICE RAN WITHOUT USER ACTIVATION WHEN THE CABLE WAS FLEXED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY, THE DEVICE CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION, AND SUBSEQUENTLY THE DEVICE RAN WITHOUT USER ACTIVATION WHEN THE CABLE WAS FLEXED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447318 | ES6 STERNUM SAW | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | HWE | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |