Report

Report Number
3004209178-2014-13734
Event Type
Injury
Date Received
July 31, 2014
Report Date
July 17, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: PRODUCT ID 97740, SERIAL # (B)(4), PRODUCT TYPE PROGRAMMER, PATIENT; PRODUCT ID 97754, SERIAL # (B)(4), PRODUCT TYPE RECHARGER; PRODUCT ID 39565-65, SERIAL # (B)(4), IMPLANTED: (B)(4) 2014, PRODUCT TYPE LEAD. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT¿S POCKET WAS SWOLLEN AND THE PATIENT HAD COMMUNICATION/TELEMETRY AND COUPLING ISSUES. IT WAS NOTED THAT MEDICAL INTERVENTION WAS NEEDED AND THE POCKET WAS ASPIRATED. IT WAS FURTHER NOTED THAT IMPEDANCE TESTING AND REPROGRAMMING WAS DONE. THE REPORTER STATED THAT THERE WAS FLUID IN THE POCKET THAT IMPEDED COMMUNICATION WITH THE PATIENT PROGRAMMER AND COUPLING WITH THE RECHARGER. THE REPORTER FURTHER STATED THE PATIENT WAS HAVING A DIFFICULT TIME CHARGING DUE TO SWELLING IN THE IMPLANTABLE NEUROSTIMULATOR (INS) POCKET. IT WAS NOTED THAT THE CHARGING LOG CONFIRMED AN AVERAGE OF FOUR COUPLING BARS FOR THE LAST FEW RECHARGES. IT WAS FURTHER NOTED THE PATIENT¿S HEALTHCARE PROFESSIONAL ASPIRATED THE INS POCKET AND 20 ML OF BLOODY FLUID WAS ASPIRATED. THE REPORTER STATED THE PATIENT WAS THEN ABLE TO GET EIGHT COUPLING BARS WITHOUT ISSUE. IT WAS NOTED THE PATIENT HAD PAIN AT THE IMPLANT SITE. IT WAS FURTHER NOTED THE PATIENT STATUS AT THE TIME OF THIS REPORT WAS ALIVE WITH NO INJURY. IT WAS NOTED THE PATIENT WAS FEELING MUCH BETTER AND THEIR COVERAGE WAS VERY GOOD. ADDITIONAL INFORMATION RECEIVED REPORTED THE PATIENT STARTED HAVING COUPLING AND RECHARGING ISSUES SINCE THEIR INS POCKET REVISION ON (B)(6) 2014. ADDITIONAL INFORMATION WAS REQUESTED, BUT WAS NOT AVAILABLE AS OF THE DATE OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
446670 SURESCAN STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW MEDTRONIC MED REL MEDTRONIC PUERTO RICO 97714

Patients

Seq Age Sex Outcome Treatment
1 00044 YR Required Intervention