FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3970051 · Received July 31, 2014

Report

Report Number
3004209178-2014-88357
Event Type
Malfunction
Date Received
July 31, 2014
Date of Event
July 1, 2014
Report Date
July 1, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD A MOTOR ERROR ALARM DURING REWIND DUE TO CORRODED MOTOR HOME SWITCH. INSULIN PUMP WAS UNABLE TO PERFORM DISPLACEMENT TEST AND VERIFY THE MOTION SENSOR TEST ALARM, MOTOR POSITION ENCODER ERROR ALARM DUE TO MOTOR ERROR ALARM. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MULTIPLE ERRORS, INCLUDING MOTOR ERROR. THE BATTERY WAS CHANGED. THE CUSTOMER'S BLOOD GLUCOSE WAS 219 MG/DL. NOTHING FURTHER WAS REPORTED.