PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-88354
- Event Type
- Malfunction
- Date Received
- July 31, 2014
- Date of Event
- July 1, 2014
- Report Date
- July 1, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
NO BUTTON ERROR ALARM OR UNEXPECTED AUDIO ANOMALY WAS NOTED. THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO WEAK BATTERY ALARM COULD BE TESTED DUE TO THE UNRESPONSIVE BUTTONS. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS, SCRATCHED RESERVOIR TUBE WINDOW, CRACKED RESERVOIR TUBE, CRACKED RESERVOIR TUBE LIP AND A MISSING END CAP STICKER.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE DEVICE KEPT BEEPING AND IT WOULD NOT ALLOW THE CUSTOMER TO PRESS ANY BUTTONS OR CLEAR THE ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 9.1 MMOL/L. IT WAS REPORTED THAT PRIOR TO THE BUTTON ERROR ALARM THE INSULIN PUMP ALSO ALARMED WEAK BATTERY. THE CUSTOMER CHANGED THE BATTERY WHICH TRIGGERED THE BUTTON ERROR ALARM. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 446598 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754CAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |