Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, ¿MATERIAL SENSITIVITY REACTIONS.¿ NUMBER 6 STATES, ¿INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS.¿ NUMBER 14 STATES, ¿INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN.¿ NUMBER 15 STATES, ¿ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES.¿ THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-06540 & 06541).

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.

Description of Event or Problem · 1

LEGAL COUNSEL FOR PATIENT REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6), 2006. PATIENT'S LEGAL COUNSEL FURTHER REPORTED PATIENT ALLEGATIONS OF PAIN, DAMAGE TO BONE/TISSUE, LACK OF MOBILITY, METAL POISONING, METALLOSIS AND ELEVATED METAL ION LEVELS. SUBSEQUENTLY, PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6), 2012. A REVIEW OF THE INVOICE HISTORY CONFIRMED THE SURGERY DATES AND INDICATES THE MODULAR HEAD, TAPER ADAPTER AND ACETABULAR CUP WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Description of Event or Problem · 1

LEGAL COUNSEL FOR PATIENT REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2006. PATIENT'S LEGAL COUNSEL FURTHER REPORTED PATIENT ALLEGATIONS OF PAIN, DAMAGE TO BONE/TISSUE, LACK OF MOBILITY, METAL POISONING, METALLOSIS AND ELEVATED METAL ION LEVELS. SUBSEQUENTLY, PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2012. A REVIEW OF THE INVOICE HISTORY CONFIRMED THE SURGERY DATES AND INDICATES THE MODULAR HEAD, TAPER ADAPTER AND ACETABULAR CUP WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. ADDITIONAL INFORMATION RECEIVED IN PATIENT'S MEDICAL RECORDS REVEALED THE (B)(6) 2012 REVISION WAS DUE TO A GRINDING NOISE. THE PATIENT'S OPERATIVE REPORT NOTED GRAY, THIN, WATER FLUID WITH METALLIC PARTICLES; GRAY MATTER, AND METAL DEBRIS. FURTHER INFORMATION IN PATIENT'S OPERATIVE REPORT NOTED COMPLICATION OF A GREATER TROCHANTERIC FRACTURE, WHICH WAS REPAIRED AT THE END OF THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
441523 M2A-MAGNUM PF CUP 56ODX50ID PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 147800

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization| R