M2A-MAGNUM MOD HD SZ 52MM
Report
- Report Number
- 0001825034-2014-06535
- Event Type
- Injury
- Date Received
- July 29, 2014
- Date of Event
- June 12, 2008
- Report Date
- July 7, 2014
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWA
- PMA / PMN Number
- PK042037
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- ATTORNEY
Narratives
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, ¿MATERIAL SENSITIVITY REACTIONS.¿ NUMBER 6 STATES, ¿INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS.¿ NUMBER 14 STATES, ¿INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN.¿ NUMBER 15 STATES, ¿ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES.¿ THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 2 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-06534 & 06535).
LEGAL COUNSEL FOR PATIENT REPORTED PATIENT UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2006. PATIENT'S LEGAL COUNSEL FURTHER REPORTED PATIENT ALLEGATIONS OF PAIN, DAMAGE TO BONE/TISSUE, LACK OF MOBILITY AND ELEVATED METAL ION LEVELS. SUBSEQUENTLY, PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2008. DURING THE PROCEDURE, THE MODULAR HEAD AND ACETABULAR CUP WAS REMOVED AND REPLACED. A REVIEW OF THE INVOICE HISTORY CONFIRMED THE SURGERY DATES AND WHICH COMPONENTS WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 440781 | M2A-MAGNUM MOD HD SZ 52MM | PROSTHESIS, HIP | KWA | BIOMET ORTHOPEDICS | N/A | 276090 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |