FDA Adverse Event Malfunction Summary report: N

ROTATABLE SNARE

MDR report key: 3964833 · Received July 29, 2014

Report

Report Number
3005099803-2014-02592
Event Type
Malfunction
Date Received
July 29, 2014
Date of Event
June 2, 2014
Report Date
July 14, 2014
Manufacturer
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
Product Code
FDI
PMA / PMN Number
K992477
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FINDINGS: CATHETER DETACHED. VISUAL EVALUATION OF THE RETURNED DEVICE FOUND THAT THE HANDLE CANNULA WAS SLIGHTLY BENT, THE FLARE WAS DETACHED FROM THE CATHETER, AND THE CATHETER IS KINKED. THE CATHETER HAD MARKS FROM THE FLARE PROCESS PERFORMED DURING MANUFACTURING. FUNCTIONAL EVALUATION COULD NOT BE PERFORMED DUE TO THE CONDITION OF THE RETURNED DEVICE. THE COMPLAINT WAS CONFIRMED; THE DETACHED FLARE PREVENTED THE SNARE FROM EXTENDING OUT OF THE SHEATH. HOWEVER, REVIEW AND ANALYSIS OF ALL AVAILABLE INFORMATION FAILED TO INDICATE A ROOT CAUSE OR PROBABLE ROOT CAUSE FOR THIS EVENT. THERE IS AN INVESTIGATION IN PLACE TO ADDRESS THIS ISSUE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PRODUCT SPECIFICATIONS AT THE TIME OF RELEASE TO DISTRIBUTION.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ROTATABLE SMALL OVAL SNARE WAS UNPACKED DURING A PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING FUNCTIONAL TESTING, THE SNARE LOOP FAILED TO EXTEND OUT OF THE SHEATH. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. THIS EVENT HAS BEEN DEEMED REPORTABLE BASED ON THE INVESTIGATION RESULTS: FLARE DETACHED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
442248 ROTATABLE SNARE SNARE, FLEXIBLE FDI BOSTON SCIENTIFIC - COSTA RICA (COYOL) M00561821 0016744780

Patients

Seq Age Sex Outcome Treatment
1