FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3964785
·
Received June 19, 2014
Report
- Report Number
- 1720753-2014-05205
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- June 3, 2014
- Report Date
- June 18, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE POWER MOTOR RELAY PCB CONNECTIONS WERE EVALUATED AND RESEATED. A COMPLETE GENERATOR CALIBRATION WAS PERFORMED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE FSE REPORTED THAT THE SYSTEM HAD INTERMITTENT INTERLOCK ERRORS. THIS CAUSES THE SYSTEM TO SHUT DOWN, RESULTING IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360896 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |