SOVEREIGN COMPACT CONSOLE
Report
- Report Number
- 3006695864-2014-00363
- Event Type
- Malfunction
- Date Received
- July 29, 2014
- Date of Event
- July 11, 2014
- Report Date
- July 11, 2014
- Manufacturer
- ABBOTT MEDICAL OPTICS
- Product Code
- HQC
- PMA / PMN Number
- K060366
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
Narratives
NOT PROVIDED, AS TO NO PATIENT INJURY IS REPORTED. THE CATEGORIES FOR OUTCOMES ATTRIBUTED TO ADVERSE EVENTS IS NOT APPLICABLE AS TO PRODUCT PROBLEM SELECTED. THE PHACOEMULSIFICATION MACHINE WAS EXAMINED AND TESTED AT THE CUSTOMER LOCATION BY AN ABBOTT FIELD SERVICE SPECIALIST (FSS). THE FSS VISUALLY INSPECTED THE UNIT AND FOUND BURNT COMPONENTS ON THE SUBSYSTEM CONTROLLER (SSC) PRINTED CIRCUIT BOARD ASSEMBLY (PCBA) AND THE SINGLE BOARD COMPUTER (SBC) PRINTED CIRCUIT BOARD (PCB). THE FSS RECTIFY THE REPORTED ISSUE BY REPLACING THE SSC PCBA AND THE SBC PCB. THE FSS TESTED FOR ALL MODES OF OPERATION. PERFORM A FIELD SERVICE CHECKLIST. THE MACHINE WAS FOUND TO MEET AMO SPECIFICATIONS. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED. PLACEHOLDER.
IN INITIAL REPORT, MODEL NUMBER IS INCORRECT. THE CORRECT MODEL NUMBER HAS BEEN PROVIDED. CMP680300. (B)(6). (B)(4). ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.
THE CLINIC REPORTED AFTER THE COMPLETION OF CATARACT EXTRACTION PROCEDURE, THE COMPACT PHACOEMULSIFICATION UNIT HAD TURNED OFF AND WOULD NOT POWER ON. FURTHERMORE, THE CLINIC REPORTED VISIBLE SMOKE WITH AN ODOR EMITTING FROM THE INSIDE OF THE UNIT. THE CLINIC REPORTED NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 441575 | SOVEREIGN COMPACT CONSOLE | SOVEREIGN COMPACT | HQC | ABBOTT MEDICAL OPTICS | CMP680300 |