ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-21722
- Event Type
- Malfunction
- Date Received
- July 29, 2014
- Report Date
- July 18, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/20/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT UNEXPLAINABLE POWER REBOOTS HAD OCCURRED. VISUAL INSPECTION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED BELOW THE GRIP PAD. THERE WAS EVIDENCE OF MOISTURE OBSERVED INSIDE THE BATTERY COMPARTMENT. NO BATTERY CAP WAS RETURNED WITH THE PUMP. A TEST BATTERY CAP WAS ABLE TO BE FULLY SECURED TO THE PUMP AND WAS USED TO COMPLETE ALL TESTING. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER REBOOTS, POWER LOSSES, OR CALL SERVICE ALARMS OCCURRING. A LEAK TEST WAS PERFORMED AND NO LEAKS WERE FOUND. THE PUMP CASE WAS REMOVED AND NO EVIDENCE OF MOISTURE CONTAMINATION WAS FOUND INSIDE THE PUMP. THE COMPLAINT THAT THE PUMP HAD NO POWER WAS OBSERVED IN THE PUMP HISTORY BUT WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. THE REPORTER ALLEGED THAT THE PUMP HAD NO POWER. IT WAS NOTED THAT THERE WAS MOISTURE/CORROSION IN THE BATTERY COMPARTMENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 441877 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |