SYMBIQ DCHANNEL 3.01
Report
- Report Number
- 9615050-2014-03765
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Date of Event
- January 1, 2011
- Report Date
- March 21, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042800
- Removal / Correction Number
- FA305-02
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE DEVICE PASSED TESTING. NO E630 (SCREW ROTATION ERROR) ERROR CODE OCCURRED DURING TESTING; HOWEVER AN E630 (SCREW ROTATION ERROR) ERROR CODE WAS NOTED IN THE DEVICE HISTORY BUT WAS NOT DUPLICATED DURING TESTING. ALTHOUGH THE DEVICE PASSED TESTING, THE DEVICE HAS BEEN IDENTIFIED A PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE DEVICE MECHANISM WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER OF A S320 (PLUNGER HOMING TIMEOUTS) MALFUNCTION ALARM COE. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, THE DEVICE ALARMED WITH AN S31 (MOTOR ERROR-PMC, LEFT) MALFUNCTION ALARM CODE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 318149 | SYMBIQ DCHANNEL 3.01 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |