UNKNOWN DEPUY HIP LINER
Report
- Report Number
- 1818910-2014-24351
- Event Type
- Injury
- Date Received
- July 29, 2014
- Date of Event
- July 8, 2014
- Report Date
- July 8, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- KWA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE PATIENT WAS REVISED BECAUSE OF DISLOCATION. DOI: APPROX. 5 YRS DOR: (B)(6) 2014 (RIGHT HIP). THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODES REQUIRED WERE NOT PROVIDED. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH (B)(4) APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THE PATIENT WAS REVISED BECAUSE OF DISLOCATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 440978 | UNKNOWN DEPUY HIP LINER | HIP ACETABULAR INSERT/LINER | KWA | DEPUY ORTHOPAEDICS, INC. | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |