FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM PEDIATRICS CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 3964512 · Received July 29, 2014

Report

Report Number
3004753838-2014-25074
Event Type
Malfunction
Date Received
July 29, 2014
Date of Event
June 30, 2014
Report Date
July 1, 2014
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE RETURNED DEVICE DID NOT CONFIRM THE REPORTED INTERMITTENT AUDIO OUTPUT.

Description of Event or Problem · 1

PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM INTERMITTENT AUDIO OUTPUT PATIENT EXPERIENCED ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, DURING THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT'S MOTHER TO TEST THE ALERT FUNCTIONALITY AND REPORTED ALERTS WERE FUNCTIONAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
440894 G4 PLATINUM PEDIATRICS CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT22430

Patients

Seq Age Sex Outcome Treatment
1 9 YR