FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3964495
·
Received May 23, 2014
Report
- Report Number
- 3008642652-2014-01577
- Event Type
- Malfunction
- Date Received
- May 23, 2014
- Date of Event
- April 22, 2014
- Report Date
- May 22, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF ELECTRODE BELT SN: (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK TE PAD MESSAGES) WAS CONFIRMED. AS RECEIVED, THE BELT FAILED INCOMING TESTING. UPON EVALUATION, THE REAR PULSE WIRE WAS OPEN INSIDE THE TRUNK CABLE. THE CAUSE OF THE CHECK TE PAD MESSAGES IS THE OPEN PULSE WIRE. THE ROOT CAUSE OF THE OPEN WIRE CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE OPEN WIRE. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) FEMALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CONSTANT CHECK THERAPY PAD ALARMS. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307388 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |