FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3964495 · Received May 23, 2014

Report

Report Number
3008642652-2014-01577
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
April 22, 2014
Report Date
May 22, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION OF ELECTRODE BELT SN: (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK TE PAD MESSAGES) WAS CONFIRMED. AS RECEIVED, THE BELT FAILED INCOMING TESTING. UPON EVALUATION, THE REAR PULSE WIRE WAS OPEN INSIDE THE TRUNK CABLE. THE CAUSE OF THE CHECK TE PAD MESSAGES IS THE OPEN PULSE WIRE. THE ROOT CAUSE OF THE OPEN WIRE CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE OPEN WIRE. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CONSTANT CHECK THERAPY PAD ALARMS. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
307388 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 56 YR