MDR report key: 3964446 · Received June 19, 2014

Report

Report Number
1720753-2014-05199
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
June 3, 2014
Report Date
June 18, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE MOTOR CONTROL PCB WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS A STATOR NOT ON ERROR MESSAGE. THIS ERROR CAUSES THE SYSTEM TO SHUT DOWN, RESULTING IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.