FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3964204
·
Received July 29, 2014
Report
- Report Number
- 1416980-2014-24533
- Event Type
- Malfunction
- Date Received
- July 29, 2014
- Date of Event
- July 4, 2014
- Report Date
- July 4, 2014
- Manufacturer
- SHARP CORPORATION OSAKA JAPAN
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. VISUAL INSPECTION, FUNCTIONAL TESTING, AND BATTERY TESTING WERE PERFORMED. THE DAMAGED BATTERY WAS IDENTIFIED DURING THE VISUAL INSPECTION. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED. TO CORRECT THE CONDITION, THE BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING PRODUCT EVALUATION, A FLOGARD INFUSION PUMP WAS FOUND WITH A DAMAGED BATTERY. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 442532 | FLOGARD | PUMP, INFUSION | FRN | SHARP CORPORATION OSAKA JAPAN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |