FDA Adverse Event Malfunction Summary report: N

SYMBIQ 3.13 DUAL CHA

MDR report key: 3964176 · Received May 28, 2014

Report

Report Number
9615050-2014-03675
Event Type
Malfunction
Date Received
May 28, 2014
Date of Event
January 1, 2014
Report Date
April 29, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING TESTING AT THE USER FACILITY, AT START UP THE DEVICE ALARMED WITH A S321 (MOTOR ERROR-PMC, LEFT) AND A S421 (MOTOR ERROR-PMC, RIGHT) MALFUNCTION ALARM CODE. PRIOR TO TESTING, THE DEVICE HAD BEEN RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
314886 SYMBIQ 3.13 DUAL CHA 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK