FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ DUALCHAN INF
MDR report key: 3964159
·
Received May 28, 2014
Report
- Report Number
- 9615050-2014-03681
- Event Type
- Malfunction
- Date Received
- May 28, 2014
- Date of Event
- January 1, 2014
- Report Date
- April 28, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0070-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, THE DEVICE TOUCHSCREEN WAS FOUND TO BE OUT OF CALIBRATION. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
DURING TESTING AT THE USER FACILITY, IT WAS NOTED THE TOUCHSCREEN WAS OUT OF CALIBRATION. PRIOR TO TESTING, THE DEVICE HAD BEEN RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314906 | SYMBIQ DUALCHAN INF | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |