FDA Adverse Event
Malfunction
Summary report: N
SCD EXPRESS W/EXT HOOK
MDR report key: 3963887
·
Received May 21, 2014
Report
- Report Number
- 3006451981-2014-00636
- Event Type
- Malfunction
- Date Received
- May 21, 2014
- Date of Event
- May 20, 2014
- Report Date
- May 21, 2014
- Manufacturer
- COVIDIEN
- Product Code
- JOW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
SUBMIT DATE: (B)(4) 2014. AN INVESTIGATION IS CURRENTLY UNDERWAY. UPON COMPLETION, THE RESULTS WILL BE FORWARDED.
Description of Event or Problem · 1
IT WAS REPORTED TO COVIDIEN ON (B)(4) 2014 THAT AN SCD CONTROLLER WAS RETURNED TO A LOCAL COVIDIEN SERVICE CENTER AS A BACK TO STOCK UNIT. A SERVICE TECHNICIAN FOUND THAT THE POWER CORD WAS CUT EXPOSING THE AC NEUTRAL COOPER WIRE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 301184 | SCD EXPRESS W/EXT HOOK | SCD CONTROLLER | JOW | COVIDIEN | U9525E | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |