FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3963789 · Received May 7, 2014

Report

Report Number
1720753-2014-04010
Event Type
Malfunction
Date Received
May 7, 2014
Date of Event
April 16, 2014
Report Date
May 7, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HIGH VOLTAGE TANK AND LEMO CABLE SOCKET WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO DISPLAY AN IMAGE WHILE FLUOROING. THIS RESULTED IN LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF DEATH OR SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
274330 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1