FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3963786 · Received May 7, 2014

Report

Report Number
1720753-2014-04014
Event Type
Malfunction
Date Received
May 7, 2014
Date of Event
April 21, 2014
Report Date
May 7, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SNUBBER PCB WAS EVALUATED AND REPLACED. A COMPLETE FILAMENT CALIBRATION WAS PERFORMED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING FLUORO THE LEFT (LIVE) MONITOR WAS BLACK. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT.