FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3963756 · Received May 7, 2014

Report

Report Number
1720753-2014-04005
Event Type
Malfunction
Date Received
May 7, 2014
Date of Event
April 23, 2014
Report Date
May 7, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE VIDEO CONTROL BOARD AND VIDEO WIRING HARNESS WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

PER THE FSE, THE CUSTOMER REPORTED THAT THE IMAGES WERE UNUSABLE. THIS COULD RESULT IN THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273460 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1