CONSTELLATION VISION SYSTEM
Report
- Report Number
- 2028159-2014-00829
- Event Type
- Malfunction
- Date Received
- May 7, 2014
- Date of Event
- April 7, 2014
- Report Date
- April 10, 2014
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CTR
- Product Code
- HQC
- PMA / PMN Number
- K101285
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION IS IN PROGRESS. A SAMPLE HAS BEEN RECEIVED AND IN-HOUSE TESTING IN PROGRESS. THE DEVICE HISTORY RECORDS (DHR) FOR THE LOTS WERE REVIEWED. AN ACTUATION ISSUE WAS FOUND ON ONE OF THE TWO COMPONENT LOTS, HOWEVER, THE SUSPECTED PRODUCT WAS RE-INSPECTED AND THE PRODUCT WAS RELEASED ACCORDING TO THE MFR'S ACCEPTANCE CRITERIA. A COMPLAINT HISTORY REVIEW INDICATES ZERO ADD'L COMPLAINTS WERE ASSOCIATED WITH THE ONE ACTUATION ISSUE COMPONENT LOTS. A ROOT CAUSE HAS NOT BEEN IDENTIFIED. (B)(4).
A SCRUB TECH REPORTED THAT THE PROBE WOULD NOT CUT AT HIGH SPEED, ONLY AT THE LOWER SPEED DURING VITRECTOMY PROCEDURE. SHE FURTHER CLARIFIED THAT WHEN THE SURGEON WAS IN THE PT'S EYE AND ATTEMPTED TO INCREASE CUTTER SPEED, THE CUTTER STOPPED OPENING AND CLOSING. THE SCRUB TECH COULD NOT RECALL IF ASPIRATION WAS STILL WORKING OR NOT. THE PROBE WAS REPLACED TO CONTINUE THE CASE. THERE WAS NO HARM TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274311 | CONSTELLATION VISION SYSTEM | UNIT, PHACOFRAGMENTATION | HQC | ALCON - IRVINE TECHNOLOGY CTR | B | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | CUSTOM PAK| TOTAL PLUS PAK 25+, 5000 CUTS PER MINUTE STD |