FDA Adverse Event
Malfunction
Summary report: N
TRUERESULT
MDR report key: 3963640
·
Received March 5, 2014
Report
- Report Number
- 1052693-2014-00077
- Event Type
- Malfunction
- Date Received
- March 5, 2014
- Date of Event
- February 5, 2014
- Report Date
- March 5, 2014
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
CONSUMER COMPLAINT OF ABOUT HIGH BLOOD RESULTS. CUSTOMER NORMALLY RANGES AROUND 140 MG/DL TWO HOURS AFTER MEALS. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE HIGHEST RESULT IN MEMORY (348) AND THE LOWEST NORMAL RESULT (140) IS LOCATED IN ZONE C. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 132983 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PP1406 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |