FDA Adverse Event Malfunction Summary report: N

INRATIO PT/INR TEST STRIPS

MDR report key: 3963476 · Received May 30, 2014

Report

Report Number
2027969-2014-00520
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 13, 2014
Report Date
May 19, 2014
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
K092987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT IS INDICATED THAT PRODUCT IS NOT RETURNING FOR EVALUATION. THEREFORE, INVESTIGATION OF THE COMPLAINT TO DETERMINE ROOT CAUSE CANNOT BE COMPLETED. SINCE THE PRODUCT ASSOCIATED WITH THE COMPLAINT WAS NOT RETURNED, A REVIEW OF IN-HOUSE TESTING DATA WAS PERFORMED. RETAIN STRIP TESTING RESULTS MET BOTH ACCURACY AND REPEATABILITY CRITERIA. THE PRODUCT PERFORMED AS EXPECTED AND NO PRODUCT DEFICIENCIES WERE OBSERVED. IMPROPER USE OF THE METER WAS IDENTIFIED IN THE COMPLAINT. THIS COULD NOT BE RULED OUT AS A CAUSE OF THE UNEXPECTED RESULTS. THE MANUFACTURING RECORDS FOR THE LOT WERE REVIEWED. THE LOT MET SPECIFICATIONS AND NO NON-CONFORMANCES WERE DOCUMENTED. NO FURTHER INVESTIGATION WIL BE PURSUED AT THIS TIME. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT INRATIO RESULTS. RESULTS AS FOLLOWS: DATE: (B)(6)_ 2014. LAB: 0.5. TIME BETWEEN TESTS: LESS THAN AN HOUR. THE CUSTOMER STATED THAT THE METER WAS NOT IN THE CORRECT MODE WHEN FINGER STICK WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320101 INRATIO PT/INR TEST STRIPS PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100139 340875

Patients