FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 3963377 · Received May 30, 2014

Report

Report Number
1720753-2014-04608
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 8, 2014
Report Date
May 30, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE POWER SUPPLY WAS ADJUSTED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY SHUT DOWN. THIS RESULTED IN AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THIS COULD CAUSE PROCEDURAL DELAYS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
318119 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1