FDA Adverse Event Malfunction Summary report: N

OPTIFLUX 180NRE DIALYZER FINISHED ASSY.

MDR report key: 3963362 · Received June 19, 2014

Report

Report Number
1713747-2014-00303
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
May 23, 2014
Report Date
June 4, 2014
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
FJI
PMA / PMN Number
K002277
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PLANT INVESTIGATION HAS NOT YET BEEN COMPLETED. A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

A HEMODIALYSIS INPATIENT USER FACILITY HAS REPORTED THAT DURING TREATMENT A BLOOD LEAK OCCURRED. THE LEAK WAS VISUALLY OBSERVED IN THE DIALYSATE. TEST STRIPS WERE NOT USED. THE MACHINE ALARMED. ESTIMATED BLOOD LOSS WAS 225 ML. PATIENT HAD NO ADVERSE EFFECTS AND NO MEDICAL INTERVENTION WAS REQUIRED. THE PATIENT COMPLETED TREATMENT WITH A NEW SET-UP. SAMPLE AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359174 OPTIFLUX 180NRE DIALYZER FINISHED ASSY. FJI FRESENIUS MEDICAL CARE NORTH AMERICA 14CUO2019

Patients

Seq Age Sex Outcome Treatment
1 38 YR FRESENIUS 2008K HEMODIALYSIS MACHINE