FDA Adverse Event Malfunction Summary report: N

HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC

MDR report key: 3963327 · Received May 29, 2014

Report

Report Number
1037905-2014-00222
Event Type
Malfunction
Date Received
May 29, 2014
Report Date
May 2, 2014
Manufacturer
WILSON-COOK MED INC
Product Code
KNQ
PMA / PMN Number
K090183
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OUR EVALUATION OF THE RETURNED PRODUCT CONFIRMED THE REPORT. APPROXIMATELY 38CM FROM DISTAL END IS A CRACK IN THE CATHETER. WHEN INFLATION OF THE DEVICE WAS ATTEMPTED WITH A 60CC SYRINGE AND AN INFLATION HANDLE, A STREAM OF WATER CAN BE SEEN EXITING THE CATHETER. THE DEVICE COULD NOT BE INFLATED OR HOLD PRESSURE IN THIS CONDITION. TO CHECK THE INTEGRITY OF THE BALLOON THE CATHETER WAS CUT DISTAL OF THE CRACKED IN THE CATHETER. THE BALLOON WAS INFLATED WITH A 60CC SYRINGE AND AN INFLATION HANDLE. NO LEAKS OR RUPTURES IN THE BALLOON WAS OBSERVED AND THE BALLOON HELD PRESSURE. NO PART OF THE DEVICE IS MISSING. A PRODUCT-SPECIFIC DISCREPANCY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THIS OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. THE INSTRUCTIONS FOR USE DIRECT THE USER TO APPLY NEGATIVE PRESSURE TO THE BALLOON TO FACILITATE PASSAGE THROUGH THE ENDOSCOPE. THE APPLICATION OF NEGATIVE PRESSURE WILL ASPIRATE ALL RESIDUAL AIR FROM THE BALLOON AND EASE ENDOSCOPIC ADVANCEMENT. THE INSTRUCTIONS FOR USE DIRECT THE USER TO APPLY A LUBRICATING AGENT TO THE BALLOON TO FACILITATE PASSAGE THROUGH THE ENDOSCOPE ACCESSORY CHANNEL. THIS ACTIVITY WILL AID IN ENDOSCOPIC ADVANCEMENT AND CATHETER PRESERVATION. THE INSTRUCTIONS FOR USE STATE: "VISUALLY INSPECT WITH PARTICULAR ATTENTION TO KINKS, BENDS AND BREAKS. IF AN ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITION, DO NOT USE." KINKS AND/OR BENDS AND/OR CRACKS IN THE CATHETER CAN OCCUR IF THE DEVICE EXPERIENCES EXCESSIVE PRESSURE DURING USE AND/OR GENERAL HANDLING. THE INSTRUCTIONS FOR USE FOR THIS PRODUCT LINE ADVISE THE USER TO ADVANCE THE DEVICE THROUGH THE ACCESSORY CHANNEL IN SHORT INCREMENTS. THIS ACTIVITY WILL AID IN DEVICE PRESERVATION. PRIOR TO DISTRIBUTION, ALL HERCULES 3 STAGE ESOPHAGEAL-PYLORIC-COLONIC WIREGUIDED BALLOONS ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE. SEE SCANNED PAGE.

Description of Event or Problem · 1

DURING AN ENDOSCOPY PROCEDURE, A COOK HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC WAS USED. THE BALLOON BURST DURING THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
315673 HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC KNQ, DILATOR, ESOPHAGEAL KNQ WILSON-COOK MED INC W3262732

Patients

Seq Age Sex Outcome Treatment
1 NP