PLUM A+ DRIVER NEW 8
Report
- Report Number
- 9615050-2014-04088
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- May 29, 2014
- Report Date
- May 29, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE INTERMITTENTLY ALARMED WITH AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPT WITH AN UNSIGNED NOTE THAT STATED, "E321 DURING INFUSION". NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE INTERMITTENTLY ALARMED WITH AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE. THIS ERROR OCCURS WHEN THE BATTERY CANNOT BE FULLY CHARGED WITHIN 8 HOURS. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 359571 | PLUM A+ DRIVER NEW 8 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |