FDA Adverse Event Malfunction Summary report: N

PLUM A+ DRIVER NEW 8

MDR report key: 3963325 · Received June 19, 2014

Report

Report Number
9615050-2014-04088
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
May 29, 2014
Report Date
May 29, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K042081
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE INTERMITTENTLY ALARMED WITH AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPT WITH AN UNSIGNED NOTE THAT STATED, "E321 DURING INFUSION". NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE INTERMITTENTLY ALARMED WITH AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE. THIS ERROR OCCURS WHEN THE BATTERY CANNOT BE FULLY CHARGED WITHIN 8 HOURS. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359571 PLUM A+ DRIVER NEW 8 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK PLUM A+ SOFTWARE MODULE, LIST#20791, SN (B)(4)