FDA Adverse Event Malfunction Summary report: N

SYMBIQ 3.13 SINGLE C

MDR report key: 3963298 · Received June 19, 2014

Report

Report Number
9615050-2014-04092
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
January 1, 2014
Report Date
May 28, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z0069-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION WAS PERFORMED AT THE USER FACILITY BY THE FIELD SERVICE ENGINEER. DURING TESTING THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. THIS WAS DUE TO THE TOUCHSCREEN BEING OUT OF CALIBRATION. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN WAS NOT RESPONSIVE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY THE DEVICE TOUCHSCREEN WAS OUT OF CALIBRATION. NO ADDITIONAL INFORMATION WAS PROVIDED.