SYMBIQ 3.13 SINGLE C
Report
- Report Number
- 9615050-2014-04092
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- January 1, 2014
- Report Date
- May 28, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z0069-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION WAS PERFORMED AT THE USER FACILITY BY THE FIELD SERVICE ENGINEER. DURING TESTING THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. THIS WAS DUE TO THE TOUCHSCREEN BEING OUT OF CALIBRATION. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN WAS NOT RESPONSIVE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY THE DEVICE TOUCHSCREEN WAS OUT OF CALIBRATION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 359392 | SYMBIQ 3.13 SINGLE C | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |