FDA Adverse Event Malfunction Summary report: N

SSP UNITRAY KIT HR HLA-DQB1 WITH TAQ

MDR report key: 3963276 · Received June 18, 2014

Report

Report Number
2244574-2014-00180
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
May 20, 2014
Report Date
May 20, 2014
Manufacturer
LIFE TECHNOLOGIES CORPORATION
Product Code
MZI
PMA / PMN Number
BK020068
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ACCORDING TO THE CUSTOMER REPORT, THE USER OF SSP UNITRAY KIT HR HLA-DQB1 WITH TAQ (CATALOG NUMBER 451414, LOT NUMBER 0151312016) MAY RESULT IN A MISTYPE. THE EVALUATION OF THIS DEVICE IS CURRENTLY IN PROGRESS. RESULTS ARE PENDING THE EVALUATION COMPLETION.

Description of Event or Problem · 1

A CUSTOMER REPORTED A DISCREPANT TESTING RESULT BETWEEN UNITRAY DQB1 HR (SKU 451414) AND THEIR COLLABORATOR'S SEQUENCING ASSAY. THE ORIGINAL RESULT OBTAINED BY UNITRAY DQB1 HR WAS DQB1 03:03. THE RESULT OBTAINED BY SEQUENCING WAS DQB1 03:30. REPEAT TESTING USING UNITRAY DQB1 HR RESULTED DQB1 03:30 WITH A WEAK POSITIVE IN LANE 24. THE TEST WAS REPEATED 2 MORE TIMES, RESULTING IN ONE INSTANCE OF DQB1 03:03 AND 1 OF DQB1 03:30, AGAIN WITH A VERY WEAK LANE 24 POSITIVE BAND. THIS EVENT MAY RESULT IN A MISTYPE. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
356925 SSP UNITRAY KIT HR HLA-DQB1 WITH TAQ TEST, QUALITATIVE FOR HLA, NON-DIAGNOS MZI LIFE TECHNOLOGIES CORPORATION 015 1312016

Patients

Seq Age Sex Outcome Treatment
1