FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3963267
·
Received June 19, 2014
Report
- Report Number
- 1720753-2014-05245
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- June 4, 2014
- Report Date
- June 19, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE BOARDS AND CONNECTORS WERE RESEATED AND THE SOFTWARE WAS INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT AN X-RAY DISABLED ERROR MESSAGE. THE FIELD ENGINEER NOTED A FRAME SYNC ERROR IN THE LOG, WHICH WOULD CAUSE A LOSS OF THE LIVE FLUOROSCOPY IMAGE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 359158 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |