FDA Adverse Event
Malfunction
Summary report: N
ITRAK 3500
MDR report key: 3963263
·
Received June 19, 2014
Report
- Report Number
- 1720753-2014-05234
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- June 4, 2014
- Report Date
- June 19, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- LLZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE UPS WAS RESET AND THE BATTERIES WERE CHANGED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THIS COULD CAUSE THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 359641 | ITRAK 3500 | RADIOLOGICAL IMAGE PROCESSING | LLZ | GE OEC MEDICAL SYSTEMS (SLC) | ITRAK 3500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |