FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3963240 · Received June 19, 2014

Report

Report Number
1720753-2014-05249
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
June 4, 2014
Report Date
June 19, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE DISPLAY ADAPTOR WILL BE REPLACED BY THE CUSTOMER. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM INTERMITTENTLY HAD JITTERY AND BLANKING OUT IMAGES. THIS ISSUE WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO INABILITY TO SEE THE LIVE IMAGE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359357 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1