FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3963144
·
Received June 19, 2014
Report
- Report Number
- 1720753-2014-05259
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- June 3, 2014
- Report Date
- June 19, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE WORKSTATION CONNECTORS AND PCBS WERE EVALUATED AND RESEATED. THE VOLTAGE ON C-ARM AND WORKSTATION POWER SUPPLIES WERE READJUSTED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE EXPOSURE BUTTON CONTINUED TO BEEP AND A COMMUNICATION ERROR WAS DISPLAYED. THIS COULD RESULT IN THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 359439 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |