FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3963144 · Received June 19, 2014

Report

Report Number
1720753-2014-05259
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
June 3, 2014
Report Date
June 19, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE WORKSTATION CONNECTORS AND PCBS WERE EVALUATED AND RESEATED. THE VOLTAGE ON C-ARM AND WORKSTATION POWER SUPPLIES WERE READJUSTED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE EXPOSURE BUTTON CONTINUED TO BEEP AND A COMMUNICATION ERROR WAS DISPLAYED. THIS COULD RESULT IN THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359439 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1