FDA Adverse Event Malfunction Summary report: N

ROTATING MULTIPLE CLIP APPLIER

MDR report key: 3963138 · Received July 29, 2014

Report

Report Number
3005075853-2014-05305
Event Type
Malfunction
Date Received
July 29, 2014
Date of Event
July 15, 2014
Report Date
July 18, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS FOUND THAT THE ER420 INSTRUMENT WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING THE DEVICE FOR FUNCTIONALITY THE DEVICE WAS CYCLED, FED, AND FORMED FIVE CLIPS AS INTENDED AND IT EJECTED FIVE CLIPS. FINALLY, IT LOCKED OUT AS INTENDED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A NEPHRECTOMY PROCEDURE THE DEVICE DIDN'T LOAD PROPERLY THE CLIPS. THE CLIPS WERE LOADED SIDEWAYS AND SUBSEQUENTLY EJECTED. THE CLIPS WERE MALFORMED. NO PATIENT ADVERSE CONSEQUENCES HAVE BEEN REPORTED. THE PROCEDURE WAS CARRIED OUT AND FINISHED BY USING A NEW LOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
441385 ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1