FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3963028 · Received June 23, 2014

Report

Report Number
1720753-2014-05336
Event Type
Malfunction
Date Received
June 23, 2014
Date of Event
June 6, 2014
Report Date
June 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE CINE HARD DISK DRIVE WAS EVALUATED AND IDENTIFIED AS REQUIRING REPLACEMENT. THE CUSTOMER DECLINED FURTHER SERVICE. NO CONCLUSION CAN BE DRAWN AS CONCLUSIVE REPAIR INFO IS UNAVAILABLE AT THIS TIME AND NO ADD'L SERVICE INFO WAS PROVIDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A LOSS OF VASCULAR IMAGING MODE FUNCTIONALITY. NO PT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
366073 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1