FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3962988 · Received June 11, 2014

Report

Report Number
1720753-2014-04924
Event Type
Malfunction
Date Received
June 11, 2014
Date of Event
May 22, 2014
Report Date
June 11, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FLUOROSCOPIC PEDAL WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE FE REPORTED THAT THE SYSTEM WAS NOT SAVING CINE RUNS. (DATA LOSS). THERE IS NO REPORT OF PT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
345045 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1