FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3962938
·
Received June 11, 2014
Report
- Report Number
- 1720753-2014-04948
- Event Type
- Malfunction
- Date Received
- June 11, 2014
- Date of Event
- May 23, 2014
- Report Date
- June 11, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE SYSTEM WAS TESTED, FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE FE REPORTED A 24-HOUR RECHARGE REQUIRED ERROR MESSAGE. THE FUNCTIONALITY WAS RECOVERED WITH A REBOOT. THIS RESULTED IN AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 345027 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |