FDA Adverse Event
Malfunction
Summary report: N
XMAX MOTOR
MDR report key: 3962866
·
Received June 13, 2014
Report
- Report Number
- 1045834-2013-16468
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Date of Event
- June 20, 2012
- Report Date
- June 20, 2012
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- ERL
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOLS. THE DEVICE WAS EVALUATED AND THE ISSUE OF CANNOT INSERT CUTTER WAS FOUND. THE DEVICE FAILED PRE-REPAIR DIAGNOSTIC ASSESSMENT CUTTER INSERTION VERIFICATION. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Description of Event or Problem · 1
REPORT RECEIVED FROM THE USA STATING THAT SERVICE WAS REQUIRED FOR THE DEVICE. DURING SERVICE, THE ISSUE OF CANNOT INSERT CUTTER WAS FOUND. IT IS UNKNOWN IF THE DEVICE WAS BEING USED DURING SURGERY. IT IS UNKNOWN IF INJURY OR MEDICAL INTERVENTION OCCURRED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350589 | XMAX MOTOR | ATTACHMENT | ERL | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |