ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-21647
- Event Type
- Malfunction
- Date Received
- July 29, 2014
- Report Date
- July 22, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 09/09/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE DISPLAY LENS DID NOT HAVE A CLOUDY APPEARANCE, AND NO EVIDENCE OF MOISTURE INGRESS WAS OBSERVED BEHIND THE DISPLAY LENS; THE REPORTED EVENT WAS NOT DUPLICATED. THE PUMP CASE WAS CRACKED FROM THE DISPLAY LENS TO THE CASE SEAL, AND THE PUMP FAILED A LEAK TEST DUE TO THIS DAMAGE. THE PUMP CASE WAS REMOVED, AND CORROSION WAS FOUND ON AN INTERNAL SURFACE NEAR THE LOCATION OF THE CASE DAMAGE. THE RELATION OF THE MOISTURE-INGRESS EVENT TO THE REPORTED CLOUDY-DISPLAY EVENT COULD NOT BE DETERMINED. UNRELATED TO THE REPORTED EVENT, THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (CLOUDY W/ MOISTURE) ISSUE. THE REPORTER ALLEGED THAT MOISTURE WAS EVIDENT BEHIND THE DISPLAY LENS. REPORTEDLY, THE USER WAS ABLE TO READ THE DISPLAY SCREEN SAFELY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 442193 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 6 YR |