FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3962806 · Received June 23, 2014

Report

Report Number
1720753-2014-05303
Event Type
Malfunction
Date Received
June 23, 2014
Date of Event
June 6, 2014
Report Date
June 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GPOS PCB WAS EVALUATED AND REPLACED. THE SYSTEM SOFTWARE AND CALIBRATION FILES WERE RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM ONLY BOOTED TO 5 ARROWS (NO BOOT). THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THIS COULD CAUSE THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
366085 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1