FDA Adverse Event Malfunction Summary report: N

SYMBIQ SINGLE CHANNE

MDR report key: 3962795 · Received June 13, 2014

Report

Report Number
9615050-2014-03996
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
January 1, 2014
Report Date
May 16, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE MECHANISM HAS BEEN RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED AT START UP THE DEVICE ALARMED WITH A S321 (MOTOR ERROR - PMC, LEFT) MALFUNCTION ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPT FOR AN UNSPECIFIED REASON. NO SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE PROVIDED. THERE WERE NO REPORTS OF ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPY WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350592 SYMBIQ SINGLE CHANNE 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK