FDA Adverse Event Malfunction Summary report: N

SYMBIQ SCHANNEL 3.01

MDR report key: 3962794 · Received June 13, 2014

Report

Report Number
9615050-2014-03997
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
May 16, 2014
Report Date
May 16, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
FA212-03
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE PASSED TESTING. ALTHOUGH THE DEVICE PASSED TESTING, A REVIEW OF THE HISTORY INDICATED A WHITESCREEN ERROR. THE WHITESCREEN ERROR OCCURRED WHEN THE USER TITRATES ONLY THE RATE WHEN THE DEVICE WAS DELIVERING IN THE KVO MODE. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED A WHITESCREEN ERROR. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPT FOR AN UNK REASON. NO SPECIFIC INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE PROVIDED. THERE WERE NO REPORTS OF ADVERSE PT EVENTS OR DELAYS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH A WHITESCREEN ERROR. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350788 SYMBIQ SCHANNEL 3.01 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK