FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3962768 · Received June 23, 2014

Report

Report Number
8020893-2014-01526
Event Type
Malfunction
Date Received
June 23, 2014
Date of Event
May 23, 2014
Report Date
June 2, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCO
Product Code
CBK
PMA / PMN Number
K970460
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB). THE CSE UPGRADED THE SOFTWARE TO THE CURRENT REVISION AND PERFORMED THE EXTENDED SELF-TEST (EST), SHORT SELF-TEST (SST) AND THE PERFORMANCE VERIFICATION TEST (PVT); THE DEVICE OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS. COVIDIEN REFERENCE: (B)(4).

Description of Event or Problem · 1

A REPORT RECEIVED FROM (B)(6) STATED AN 840 VENTILATOR FAILED THE POWER ON SELF TEST (POST) WHILE BEING USED ON A PT. THE VENTILATOR WAS REPORTED TO BE SET IN BILEVEL MODE. THERE WAS NO REPORT OF PT HARM AS A RESULT OF THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
364613 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCO 840

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention