FDA Adverse Event Malfunction Summary report: N

MILLENNIUM INFANT VENTILATOR

MDR report key: 3962700 · Received June 19, 2014

Report

Report Number
2020676-2014-00023
Event Type
Malfunction
Date Received
June 19, 2014
Date of Event
May 16, 2014
Report Date
June 19, 2014
Manufacturer
SECHRIST INDUSTRIES, INC.
Product Code
CBK
PMA / PMN Number
K993167
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED BY THE MANUFACTURER FOR EVAL. UPON RECEIPT, IT WAS NOTED THAT PLUG FITTING WAS MISSING FROM TH SIDE OF THE VENTILATOR. AN EVAL PLUG WAS USED, HOWEVER, THE EVALUATOR DID NOTE THAT THE THREADS THE PLUG SCREWS INTO WERE DAMAGED. EVAL OF THE VENTILATOR DID VERIFY THAT THERE WAS NO OUTPUT PRESSURE. THE DUMP VALVE WAS STUCK IN THE OPEN POSITION DIRECTLY CAUSING THE FAILURE. AS THE VENTILATOR TRIED TO BUILD PRESSURE THE PRESSURE WAS BEING RELEASED. FURTHER REVIEW OF THE DEVICE FOUND THE FOLLOWING: FOREIGN HARDWARE WAS INSTALLED IN THE VENTILATOR. THE MANUFACTURER USES STAINLESS STEEL SCREWS HOWEVER, CADMIUM PLATED SCREWS WERE FOUND. ONE OF THE TWO POPPETS WAS BEHIND THE ELECTRONIC SIDE OF THE BACK PANEL. THE POPPETS MUST BE IN FRONT OF THE PANEL TO FUNCTION. THE DIAPHRAGM AND EXHALATION BLOCK WERE NOT INSTALLED PROPERLY AT TIME OF ARRIVAL. TUBING FROM INSIDE THE VENTILATOR WAS PINCHED ON THE SENSE LINE UNDER THE BACK PLATE. THE ROOT CAUSE FOR THE DEVICE NOT FUNCTIONING PER SPECIFICATION WAS TAMPERING AND/OR ATTEMPTED MAINTENANCE. THERE IS NO SVC HISTORY FOR THIS DEVICE ON FILE WITH THE MANUFACTURER. VENTILATOR SN (B)(4) IS 6 YEARS PAST DUE FOR THE RECOMMENDED 2 YEAR OVERHAUL.

Description of Event or Problem · 1

DURING PRE-CLINICAL TESTING, THE VENTILATOR WAS NOT BUILDING PRESSURE. PER THE INFO PROVIDED, THERE WAS NO PT INVOLVEMENT AND THE DEVICE WAS REMOVED FROM SVC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359161 MILLENNIUM INFANT VENTILATOR VENTILATOR, CONTINUOUS, FACILITY USE CBK SECHRIST INDUSTRIES, INC. 23065-1

Patients

Seq Age Sex Outcome Treatment
1