INRATIO PT/INR TEST STRIPS
Report
- Report Number
- 2027969-2014-00596
- Event Type
- Malfunction
- Date Received
- June 19, 2014
- Date of Event
- May 27, 2014
- Report Date
- May 27, 2014
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K092987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NOT APPLICABLE
Narratives
IT IS INDICATED THAT PRODUCT IS NOT RETURNED FOR EVALUATION. THEREFORE, INVESTIGATION OF THE COMPLAINT TO DETERMINE ROOT CAUSE CANNOT BE COMPLETED. SINCE THE PRODUCT ASSOCIATED WITH THE COMPLAINT WAS NOT RETURNED, A REVIEW OF IN-HOUSE TESTING DATA WAS PERFORMED. RETAIN STRIPS TESTING RESULTS MET BOTH ACCURACY AND REPEATABILITY CRITERIA. THE PRODUCT PERFORMED AS EXPECTED AND NO PRODUCT DEFICIENCIES WERE OBSERVED. ALTHOUGH THE ROOT CAUSE ANALYSIS DID NOT INDICATE RETURN TESTING, IMPROPER TECHNIQUES WERE IDENTIFIED IN THE COMPLAINT. THESE COULD NOT BE RULED OUT AS A CAUSE OF THE UNEXPECTED RESULTS. THE MANUFACTURING RECORDS FOR THE LOT WERE REVIEWED. THE LOT MET SPECIFICATIONS AND NO NON-CONFORMANCES WERE DOCUMENTED. NO FURTHER INVESTIGATION WILL BE PURSUED AT THIS TIME. BASED ON THE INFO AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
"CALLER ALLEGED IMPRECISION WITH INRATIO METER. RESULTS AS FOLLOWS.: 3.6 AND 1.3. 10-15 MIN BETWEEN INRATIO TESTS. THERAPEUTIC RANGE: 2.0 -2.5. THE SAMPLE NOT APPLIED IMMEDIATELY AFTER FINGERSTICK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 359416 | INRATIO PT/INR TEST STRIPS | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | 339256 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | WARFARIN| THYROID MEDICATION - NO RECENT CHANGES. |