FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3962626 · Received June 18, 2014

Report

Report Number
8020893-2014-01470
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
January 1, 2014
Report Date
May 23, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

A REPORT RECEIVED FROM THE USA STATED AN 840 VENTILATOR EXPERIENCED A BLANK DISPLAY. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
356853 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment
1