FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3962626
·
Received June 18, 2014
Report
- Report Number
- 8020893-2014-01470
- Event Type
- Malfunction
- Date Received
- June 18, 2014
- Date of Event
- January 1, 2014
- Report Date
- May 23, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
A REPORT RECEIVED FROM THE USA STATED AN 840 VENTILATOR EXPERIENCED A BLANK DISPLAY. THERE WAS NO REPORTED PATIENT INVOLVEMENT.