FDA Adverse Event Malfunction Summary report: N

HT70 PLUS VENTILATOR

MDR report key: 3962622 · Received June 18, 2014

Report

Report Number
2023050-2014-00248
Event Type
Malfunction
Date Received
June 18, 2014
Date of Event
May 22, 2014
Report Date
May 23, 2014
Manufacturer
NEWPORT MEDICAL INSTRUMENTS, INC.
Product Code
CBK
PMA / PMN Number
K111146
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE UNIT, AND VERIFIED THE REPORTED MALFUNCTION. THE CSE REPLACED THE COIN BATTERY, WHICH RESOLVED THE REPORTED ISSUE. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MANUFACTURING SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY A CUSTOMER IN THE USA THAT, AN HT70 PLUS VENTILATOR TOUCH SCREEN WAS UNRESPONSIVE AND DISPLAYING AN ERROR CODE "COIN BATTERY NEEDS TO BE REPLACED", WHILE THE CUSTOMER WAS TRYING TO POWER ON THE UNIT. THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
356899 HT70 PLUS VENTILATOR CBK, NOU CBK NEWPORT MEDICAL INSTRUMENTS, INC. HT70

Patients

Seq Age Sex Outcome Treatment
1